devices.cryomagnetics:CS04
Cryomagnetics CS04 magnet power supply.
Usage
devices.cryomagnetics:CS04(
interface,
**kwargs,
)Typically connected via GPIB::<address>::INSTR The user shall set max_field to a reasonable value upon initialization.
Parameters
interface: str-
VISA resource name
**kwargs-
Additional arguments passed to VisaDevice. Includes the optional
max_field(float), the maximum allowed field in Tesla (defaults to 0.1 T).
Methods
| Name | Description |
|---|---|
| get_field() | Get the current magnetic field. |
| ident() | Get the device identification string. |
| reset() | Reset the magnet to zero field. |
| set_field() | Set the magnetic field to a specified value. |
| wait_field() | Wait until the magnet field reaches the specified setpoint. |
get_field()
Get the current magnetic field.
Usage
get_field()Returns
float- Current magnetic field in Tesla
ident()
Get the device identification string.
Usage
ident()Returns
str- Device identification string
reset()
Reset the magnet to zero field.
Usage
reset()Sets the upper and lower limits to zero and initiates a slow sweep to zero field.
set_field()
Set the magnetic field to a specified value.
Usage
set_field(field)Parameters
field: float- Target magnetic field in Tesla
Notes
If field is zero, it will call reset(). If field exceeds max_field, the command will be rejected. For positive fields, it sets upper limit and sweeps up. For negative fields, it sets lower limit and sweeps down.
wait_field()
Wait until the magnet field reaches the specified setpoint.
Usage
wait_field(
setpoint=None,
delta=0.0002,
)Parameters
setpoint: float = None-
Target field in Tesla. If None, uses the internal setpoint.
delta: float = 0.0002- Acceptable difference between current and target field, defaults to 0.0002 T
Notes
The method will wait until the field is within delta of the setpoint for at least two consecutive readings, then pause the sweep.