devices.keysight:PSG8257D
Keysight PSG 8257D-521 microwave signal generator.
Usage
devices.keysight:PSG8257D(
interface,
reset=True,
timeout=10000.0,
**kwargs,
)Methods
| Name | Description |
|---|---|
| __init__() | Initialize a PSG8257D device. |
| configureAmpMod() | Configure the PSG amplitude modulation (AmpMod). |
| configureCW() | Configure the PSG to output a wave with a constant frequency. |
| configurePulseMod() | Configure the pulse modulation of the output signal. |
| configureSweep() | Change the sweep settings in the given channel. |
| getSourcePower() | Get the PSG output power. |
| output() | Set the output state of the PSG. |
| readFreq() | Read the frequency parameter from the PSG. |
| readSweepParams() | Read the sweep parameters from the device. |
| reset() | Disable the signal modulation and turn off the output of the PSG. |
| setLFO() | Configure the PSG low frequency output (LFO). |
| setModulation() | Enable or disable the signal modulation. |
| setSourcePower() | Configure the PSG output power. |
| startSweep() | Prepare the PSG for triggering a sweep. |
| stopSweep() | Turn the PSG output off. |
| trigger() | Trigger the sweep(s). |
__init__()
Initialize a PSG8257D device.
Usage
__init__(
interface,
reset=True,
timeout=10000.0,
**kwargs,
)Parameters
interface: str-
The IP address and port where the device is located. e.g. TCPIP::192.168.5.102::5025::SOCKET
reset: bool = True-
If True, the device is reset on object creation using the reset method. Default is True.
timeout: int = 10000.0-
The timeout of the ethernet connection in milliseconds. Default is 10e3 ms.
**kwargs- Keyword arguments passed to the VISAdevice constructor.
configureAmpMod()
Configure the PSG amplitude modulation (AmpMod).
Usage
configureAmpMod(
AmpMod=True,
amMode="DEEP",
ampSource="INT",
intFreq=1000.0,
intShape="SINE",
ampDepth=100,
)Parameters
AmpMod: bool = True-
If True, enable amplitude modulation. If False, disable amplitude modulation. Default is True.
amMode: str = "DEEP"-
Amplitude modulation mode, either “DEEP” or “NORM”. Default is “DEEP”.
ampSource: str = "INT"-
The source of the amplitude modulation signal, which can take the values: internal:‘INT’, internal 2:‘INT2’, external:‘EXT’, external 2:‘EXT2’. Default is “INT”.
intFreq: float = 1000.0-
Frequency of the internal oscillator in Hertz, which can take values from 0.5 Hz to 1 MHz. Default is 1e3.
intShape: str = "SINE"-
Shape of the internal oscillations, which can take the values: sine:‘SINE’, triangle:‘TRI’, square:‘SQU’, ramp:‘RAMP’, noise:‘NOIS’, dual-sine:‘DUAL’, swept-sine:‘SWEP’. Default is “SINE”.
ampDepth: int = 100- Amplitude modulation in percent, which can take values from 0 to 100%. Default is 100.
configureCW()
Configure the PSG to output a wave with a constant frequency.
Usage
configureCW(freq)Parameters
freq: int, float, or str- The frequency in Hz. Can be a number or ‘MIN’/‘MAX’.
configurePulseMod()
Configure the pulse modulation of the output signal.
Usage
configurePulseMod(
PulseMod=True,
pulseSource="INT",
pulseInput="SQU",
frequency=1000.0,
)Parameters
PulseMod: bool = True-
If True, enable pulse modulation. If False, disable pulse modulation. Default is True.
pulseSource: str = "INT"-
Source of the pulse modulation signal, which can take the values: internal:‘INT’, external:‘EXT’, scalar:‘SCAL’. Default is “INT”.
pulseInput: str = "SQU"-
Internally generated modulation input for the pulse modulation, which can take the values: square:‘SQU’, free-run:‘FRUN’, triggered:‘TRIG’, doublet:‘DOUB’, gated:‘GATE’. Default is “SQU”.
frequency: float = 1000.0- Pulse rate frequency in Hertz, which can take values from 0.1 Hz to 10 MHz. Default is 1e3.
configureSweep()
Change the sweep settings in the given channel.
Usage
configureSweep(
fStart,
fStop,
fPoints,
stepDwell,
)Frequency units are in Hz. ‘MIN’/‘MAX’ arguments can be used instead of actual numbers, and use the highest/lowest setting the device is capable of.
Parameters
fStart: int or float-
The frequency at which the sweep starts.
fStop: int or float-
The frequency at which the sweep ends.
fPoints: int or str-
The number of points per sweep. Can be an integer or ‘MIN’/‘MAX’.
stepDwell: int or float- The dwell time for a step sweep.
Notes
Configures the device for step sweeping with automatic sweep timing.
getSourcePower()
Get the PSG output power.
Usage
getSourcePower()Returns
str- Power in dBm.
output()
Set the output state of the PSG.
Usage
output(state)Parameters
state: bool- If True, enable the output. If False, disable the output.
readFreq()
Read the frequency parameter from the PSG.
Usage
readFreq()Returns
float- The output frequency in Hz.
readSweepParams()
Read the sweep parameters from the device.
Usage
readSweepParams()Frequencies are returned in Hz.
Returns
tuple-
A tuple containing (fStart, fStop) where:
- fStart (float): The frequency at which the sweep starts.
- fStop (float): The frequency at which the sweep stops.
reset()
Disable the signal modulation and turn off the output of the PSG.
Usage
reset()setLFO()
Configure the PSG low frequency output (LFO).
Usage
setLFO(
LFO=True,
source="INT",
amplitude=3,
)Parameters
LFO: bool = True-
If True, enable LFO. If False, disable LFO. Default is True.
source: str = "INT"-
The source of the low frequency output, which can take the values: internal:‘INT’, internal2:‘INT2’, function:‘FUNC’, function2:‘FUNC2’ internal & internal2: for the internal source function & function2: for an internal function generator which can be configured. Default is “INT”.
amplitude: float = 3- Peak voltage (amplitude) of the low frequency output in volts, which can take values from 0-3.5V. Default is 3.
setModulation()
Enable or disable the signal modulation.
Usage
setModulation(mod=False)Parameters
mod: bool = False- If True, enable modulation. If False, disable modulation. Default is False.
setSourcePower()
Configure the PSG output power.
Usage
setSourcePower(power)Parameters
power: float- Power in dBm (setpoint between +25 and -20 dBm).
startSweep()
Prepare the PSG for triggering a sweep.
Usage
startSweep()The PSG activates the output, disables the continuous trigger and therefore enables manual triggering. Any currently running sweeps are aborted.
stopSweep()
Turn the PSG output off.
Usage
stopSweep()trigger()
Trigger the sweep(s).
Usage
trigger()Notes
The function will wait for the sweep to complete before returning.